Development and optical characterization of a multilayered head phantom for optical brain monitoring
Roldan, M., Kyriacou, P. A.
ORCID: 0000-0002-2868-485X & Awad-Pérez, K.
ORCID: 0009-0009-0875-8709 (2026).
Development and optical characterization of a multilayered head phantom for optical brain monitoring.
Journal of Biomedical Optics, 31(09),
pp. 1-18.
doi: 10.1117/1.jbo.31.9.095003
Abstract
Significance
Optical techniques are promising tools for non-invasive cerebral monitoring; however, optical measurements acquired from the forehead can be influenced by extracranial tissues and skin pigmentation, potentially complicating interpretation of cerebral signals. Tissue-mimicking phantoms provide controlled platforms for studying these effects; however, existing head phantoms employ simplified geometries and lack the features required for cerebral photoplethysmography (PPG) investigations.
Aim
We present the development and characterization of a multilayer head phantom reproducing key anatomical, optical, and hemodynamic features relevant to optical brain monitoring while enabling future investigation of confounding factors.
Approach
The phantom incorporates a custom-made silicone brain with embedded vessels, a three-dimensional-printed skull, a vascularized scalp, and interchangeable skin layers representing different pigmentation levels, integrated with an in vitro cardiovascular system enabling selective perfusion of cerebral and extracranial compartments. The optical and colorimetric properties of the tissue-mimicking layers were assessed.ResultsThe scalp and skin layers exhibited absorption and reduced scattering coefficients within ranges reported for human tissues. Colorimetric assessment confirmed physiologically relevant pigmentation characteristics across the skin layers. Using the pale skin, the phantom generated measurable PPG signals across multiple wavelengths and source–detector separations.
Conclusions
The multilayer head phantom provides an anatomically and optically representative platform for optical brain monitoring research. Its interchangeable skin layers and independently perfusable cerebral and extracranial compartments enable future controlled investigations of skin pigmentation and extracranial hemodynamics as confounding factors in cerebral optical measurements.
| Publication Type: | Article |
|---|---|
| Additional Information: | © The Authors. Published by SPIE under the terms of the Creative Commons Attribution NonCommercial-NoDerivatives (CC-BY-NC-ND) 4.0 International License. Use, distribution, or reproduction of this work in whole or in part in any medium is permitted for noncommercial purposes only, provided the work is not modified and the original work is properly cited, including the DOI. |
| Publisher Keywords: | Head phantom, Cerebral photoplethysmography, Extracranial circulation, Skin pigmentation, Optical characterization, Skin, Brain, Tissue optics, Silicon, Head, Optical sensing, Skull, Optical properties, Mixtures, Tissues |
| Subjects: | Q Science > QP Physiology R Medicine > RC Internal medicine > RC0321 Neuroscience. Biological psychiatry. Neuropsychiatry T Technology > TK Electrical engineering. Electronics Nuclear engineering |
| Departments: | School of Science & Technology School of Science & Technology > Department of Engineering |
| SWORD Depositor: |
Available under License Creative Commons Attribution Non-commercial No Derivatives.
Download (22MB) | Preview
Export
Downloads
Downloads per month over past year
Metadata
Metadata