Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter

Phillips, J. P., Langford, R. M., Chang, S. H. , Maney, K., Kyriacou, P. A. & Jones, D. P. (2010). Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter. Neurocritical Care, 13(2), pp. 278-285. doi: 10.1007/s12028-010-9349-y

Abstract

Background
A pilot investigation was undertaken to assess the performance of a novel fiber-optic cerebral pulse oximetry system. A fiber-optic probe designed to pass through the lumen of a cranial bolt of the type used to make intracranial pressure measurements was used to obtain optical reflectance signals directly from brain tissue.

Methods
Short-duration measurements were made in six patients undergoing neurosurgery. These were followed by a longer duration measurement in a patient recovering from an intracerebral hematoma. Estimations of cerebral arterial oxygen saturation derived from a frequency domain-based algorithm are compared with simultaneous pulse oximetry (SpO2) and emoximeter (SaO2) blood samples.

Results
The short-duration measurements showed that reliable photoplethysmographic signals could be obtained from the brain tissue. In the long-duration study, the mean (±SD) difference between cerebral oxygen saturation (ScaO2) and finger SpO2 (in saturation units) was
-7.47(±3.4)%. The mean (±SD) difference between ScaO2 and blood SaO2 was -7.37(±2.8)%.

Conclusions
This pilot study demonstrated that arterial oxygen saturation may be estimated from brain tissue via a fiber-optic pulse oximeter used in conjunction with a cranial bolt. Further studies are needed to confirm the clinical utility of the technique.

Publication Type: Article
Additional Information: The final publication is available at Springer via http://dx.doi.org/10.1007/s12028-010-9349-y
Publisher Keywords: Pulse oximetry, Saturation, Cerebral, Fiber-optic, Cranial bolt
Subjects: R Medicine
T Technology > TA Engineering (General). Civil engineering (General)
Departments: School of Science & Technology
School of Science & Technology > Engineering
SWORD Depositor:
[thumbnail of CRO2010 J. P. Phillips, et al (2010)Cerebral Arterial Oxygen Saturation Measurements using a Fiber-optic..., Neurocritical Care, vol. 13 issue 2 October 2010 278-285.pdf]
Preview
Text - Accepted Version
Download (404kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login