Investigation of the human oesophagus as a new monitoring site for blood oxygen saturation

Kyriacou, P. A., Moye, A. R., Choi, D. M. , Langford, R. M. & Jones, D. P. (2001). Investigation of the human oesophagus as a new monitoring site for blood oxygen saturation. Physiological Measurement, 22(1), pp. 223-232. doi: 10.1088/0967-3334/22/1/325

Abstract

Pulse oximeter probes placed peripherally may fail to give accurate values of arterial blood oxygen saturation (SpO2) when peripheral perfusion is poor. Since central blood flow may be preferentially preserved, the oesophagus was suggested as an alternative monitoring site. A reflectance oesophageal photoplethysmographic (PPG) probe and a multiplexed data acquisition system, operating simultaneously at two wavelengths and incorporating an external three-lead electrocardiogram (ECG) reference channel, has been developed. It has been used to investigate the suitability of the oesophagus as a possible monitoring site for SpO2 in cases of compromised peripheral perfusion. Oesophageal PPG signals and standard ECG traces were obtained from 16 anaesthetized patients and displayed on a laptop computer. Measurable PPG signals with high signal-to-noise ratios at both infrared and red wavelengths were obtained from all five oesophageal depths investigated. The maximum PPG amplitude occurred at 25 cm from the upper incisors in the mid-oesophagus. The measured pulse transit times (PTTs) to the oesophagus were consistent with previous measurements at peripheral sites and had a minimum value of 67 +/- 30 ms at a depth of 30 cm. There was broad agreement between the calculated values of oesophageal SpO2 and those from a commercial finger pulse oximeter.

Publication Type: Article
Additional Information: This is an author-created, un-copyedited version of an article accepted for publication in 'Physiological Measurement'. The publisher is not responsible for any errors or omissions in this version of the manuscript or any version derived from it. The Version of Record is available online at http://dx.doi.org/10.1088/0967-3334/22/1/325.
Publisher Keywords: photoplethysmography (PPG); oesophagus; monitoring; blood oxygen saturation.
Subjects: Q Science > QM Human anatomy
T Technology > TK Electrical engineering. Electronics Nuclear engineering
Departments: Professional Services > Library Services
School of Science & Technology > Engineering
SWORD Depositor:
[thumbnail of CRO 2001 P.A. KYRIACOU, et al (2001) Investigation of the Human Oesophagus..., Physiological Measurement 22(1) 223-232.pdf]
Preview
Text - Accepted Version
Download (177kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login