City Research Online

Faster Blood Flow Rate Does Not Improve Circuit Life in Continuous Renal Replacement Therapy: A Randomized Controlled Trial

Fealy, N., Aitken, L. M., du Toit, E. , Lo, S. & Baldwin, I. (2017). Faster Blood Flow Rate Does Not Improve Circuit Life in Continuous Renal Replacement Therapy: A Randomized Controlled Trial. Critical Care Medicine, 45(10), e1018-e1025. doi: 10.1097/ccm.0000000000002568

Abstract

Objectives: To determine whether blood flow rate influences circuit life in continuous renal replacement therapy.

Design: Prospective randomized controlled trial.

Setting: Single center tertiary level ICU.

Patients: Critically ill adults requiring continuous renal replacement therapy.

Interventions: Patients were randomized to receive one of two blood flow rates: 150 or 250 mL/min.

Measurements and Main Results: The primary outcome was circuit life measured in hours. Circuit and patient data were collected until each circuit clotted or was ceased electively for nonclotting reasons. Data for clotted circuits are presented as median (interquartile range) and compared using the Mann-Whitney U test. Survival probability for clotted circuits was compared using log-rank test. Circuit clotting data were analyzed for repeated events using hazards ratio. One hundred patients were randomized with 96 completing the study (150 mL/min, n = 49; 250 mL/min, n = 47) using 462 circuits (245 run at 150 mL/min and 217 run at 250 mL/min). Median circuit life for first circuit (clotted) was similar for both groups (150 mL/min: 9.1 hr [5.5–26 hr] vs 10 hr [4.2–17 hr]; p = 0.37). Continuous renal replacement therapy using blood flow rate set at 250 mL/min was not more likely to cause clotting compared with 150 mL/min (hazards ratio, 1.00 [0.60–1.69]; p = 0.68). Gender, body mass index, weight, vascular access type, length, site, and mode of continuous renal replacement therapy or international normalized ratio had no effect on clotting risk. Continuous renal replacement therapy without anticoagulation was more likely to cause clotting compared with use of heparin strategies (hazards ratio, 1.62; p = 0.003). Longer activated partial thromboplastin time (hazards ratio, 0.98; p = 0.002) and decreased platelet count (hazards ratio, 1.19; p = 0.03) were associated with a reduced likelihood of circuit clotting.

Conclusions: There was no difference in circuit life whether using blood flow rates of 250 or 150 mL/min during continuous renal replacement therapy.

Publication Type: Article
Additional Information: This is a non-final version of an article published in final form in Fealy, N., Aitken, L. M., du Toit, E., Lo, S. & Baldwin, I. (2017). Faster Blood Flow Rate Does Not Improve Circuit Life in Continuous Renal Replacement Therapy: A Randomized Controlled Trial. Critical Care Medicine, 45(10), e1018-e1025.
Publisher Keywords: acute kidney injury; circuit life; continuous hemofiltration; continuous renal replacement therapy; critical care
Departments: School of Health & Psychological Sciences > Nursing
SWORD Depositor:
[thumbnail of Fealy et al_Circuit life_accepted for CCM.pdf]
Preview
Text - Accepted Version
Download (471kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login