Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry

Campbell, P. ORCID: 0000-0002-6019-1596, Edgar, D. F ORCID: 0000-0001-9004-264X & Shah, R. ORCID: 0000-0002-6134-0936 (2021). Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry. Clinical and Experimental Optometry, 104(5), pp. 602-610. doi: 10.1080/08164622.2021.1878831

Abstract

Background: This study investigates: agreement in intraocular pressure measurements between three tonometers and Goldmann applanation tonometry (GAT); inter-optometrist agreement for each tonometer; intra-optometrist agreement for GAT; association between central corneal thickness (CCT) and IOP measurements with each tonometer.

Methods: IOP was measured using: CT-1P Non-Contact Tonometer (NCT) (Topcon Corporation, Tokyo, Japan), Pulsair IntelliPuff (Keeler Ltd., Windsor, UK) and Icare rebound tonometer (Icare, Helsinki, Finland) by two optometrists in a random order. Two GAT readings were obtained by each optometrist in a randomised masked manner. Mean differences, and 95% limits of agreement (LoA) for each measurement were calculated. CCT was measured by CT-1P pachymeter.

Results: Forty-one participants’ IOPs were measured. Mean differences (95% LoA) between NCT, Pulsair, Icare compared to GAT for one optometrist were: 0.8 (−5.4 to 6.9) mmHg, −1.7 (−8.2 to 4.8) mmHg, −1.6 (−9.0 to 5.9) mmHg. Mean differences (95% LoA) in inter-optometrist agreement for GAT, NCT, Pulsair and Icare were: 0.3 (−6.7 to 7.3) mmHg, 0.4 (−2.1 to 2.9) mmHg, −0.9 (−3.6 to 1.9) mmHg and −0.2 (−4.9 to 4.5) mmHg, respectively. Mean differences (95% LoA) for intra-optometrist agreement for GAT were 0.2 (4.3 to −4.7) mmHg and 0.1 (3.6 to −3.9) mmHg for each optometrist, respectively. There was a weak positive association between CCT and both GAT (r2 = 0.11) and NCT (r2 = 0.12).

Conclusion: Pulsair and Icare may measure IOP lower than GAT. Mean differences for inter-optometrist agreement for all tonometers were < 1 mmHg; Pulsair showed a statistically significant difference. Intra-optometrist agreement for GAT was good. IOP measurements taken by two community optometrists are comparable using tonometers used in community practice.

Publication Type: Article
Additional Information: This is the peer reviewed version of the following article: Peter Campbell, David F Edgar & Rakhee Shah (2021) Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry, Clinical and Experimental Optometry, 104:5, 602-610, which has been published in final form at https://doi.org/10.1080/08164622.2021.1878831. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions.
Publisher Keywords: Comparison, intraocular pressure, optometry, tonometer
Subjects: R Medicine > RE Ophthalmology
Departments: School of Health & Psychological Sciences > Optometry & Visual Sciences
SWORD Depositor:
[thumbnail of Campbell FINAL FOR J CLIN OPT accepted by publisher.pdf]
Preview
Text - Accepted Version
Download (349kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login