Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry
Campbell, P. ORCID: 0000-0002-6019-1596, Edgar, D. F ORCID: 0000-0001-9004-264X & Shah, R. ORCID: 0000-0002-6134-0936 (2021). Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry. Clinical and Experimental Optometry, 104(5), pp. 602-610. doi: 10.1080/08164622.2021.1878831
Abstract
Background: This study investigates: agreement in intraocular pressure measurements between three tonometers and Goldmann applanation tonometry (GAT); inter-optometrist agreement for each tonometer; intra-optometrist agreement for GAT; association between central corneal thickness (CCT) and IOP measurements with each tonometer.
Methods: IOP was measured using: CT-1P Non-Contact Tonometer (NCT) (Topcon Corporation, Tokyo, Japan), Pulsair IntelliPuff (Keeler Ltd., Windsor, UK) and Icare rebound tonometer (Icare, Helsinki, Finland) by two optometrists in a random order. Two GAT readings were obtained by each optometrist in a randomised masked manner. Mean differences, and 95% limits of agreement (LoA) for each measurement were calculated. CCT was measured by CT-1P pachymeter.
Results: Forty-one participants’ IOPs were measured. Mean differences (95% LoA) between NCT, Pulsair, Icare compared to GAT for one optometrist were: 0.8 (−5.4 to 6.9) mmHg, −1.7 (−8.2 to 4.8) mmHg, −1.6 (−9.0 to 5.9) mmHg. Mean differences (95% LoA) in inter-optometrist agreement for GAT, NCT, Pulsair and Icare were: 0.3 (−6.7 to 7.3) mmHg, 0.4 (−2.1 to 2.9) mmHg, −0.9 (−3.6 to 1.9) mmHg and −0.2 (−4.9 to 4.5) mmHg, respectively. Mean differences (95% LoA) for intra-optometrist agreement for GAT were 0.2 (4.3 to −4.7) mmHg and 0.1 (3.6 to −3.9) mmHg for each optometrist, respectively. There was a weak positive association between CCT and both GAT (r2 = 0.11) and NCT (r2 = 0.12).
Conclusion: Pulsair and Icare may measure IOP lower than GAT. Mean differences for inter-optometrist agreement for all tonometers were < 1 mmHg; Pulsair showed a statistically significant difference. Intra-optometrist agreement for GAT was good. IOP measurements taken by two community optometrists are comparable using tonometers used in community practice.
Publication Type: | Article |
---|---|
Additional Information: | This is the peer reviewed version of the following article: Peter Campbell, David F Edgar & Rakhee Shah (2021) Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry, Clinical and Experimental Optometry, 104:5, 602-610, which has been published in final form at https://doi.org/10.1080/08164622.2021.1878831. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions. |
Publisher Keywords: | Comparison, intraocular pressure, optometry, tonometer |
Subjects: | R Medicine > RE Ophthalmology |
Departments: | School of Health & Psychological Sciences > Optometry & Visual Sciences |
SWORD Depositor: |
Download (349kB) | Preview
Export
Downloads
Downloads per month over past year