City Research Online

A new software for automated counting of glistenings in intraocular lenses in vivo

Stanojcic, N., Hull, C. ORCID: 0000-0002-2205-4443, Mangieri, E. , Little, N. & O'Brart, D. (2023). A new software for automated counting of glistenings in intraocular lenses in vivo. International Journal of Ophthalmology, 16(8), pp. 1237-1242. doi: 10.18240/ijo.2023.08.08

Abstract

AIM: To assess the performance of a bespoke software for automated counting of intraocular lens (IOL) glistenings in slit-lamp images.

METHODS: IOL glistenings from slit-lamp-derived digital images were counted manually and automatically by the bespoke software. The images of one randomly selected eye from each of 34 participants were used as a training set to determine the threshold setting that gave the best agreement between manual and automatic grading. A second set of 63 images, selected using randomised stratified sampling from 290 images, were used for software validation. The images were obtained using a previously described protocol. Software-derived automated glistenings counts were compared to manual counts produced by three ophthalmologists.

RESULTS: A threshold value of 140 was determined that minimised the total deviation in the number of glistenings for the 34 images in the training set. Using this threshold value, only slight agreement was found between automated software counts and manual expert counts for the validating set of 63 images (κ=0.104, 95%CI, 0.040-0.168). Ten images (15.9%) had glistenings counts that agreed between the software and manual counting. There were 49 images (77.8%) where the software overestimated the number of glistenings.

CONCLUSION: The low levels of agreement show between an initial release of software used to automatically count glistenings in in vivo slit-lamp images and manual counting indicates that this is a non-trivial application. Iterative improvement involving a dialogue between software developers and experienced ophthalmologists is required to optimise agreement. The results suggest that validation of software is necessary for studies involving semi-automatic evaluation of glistenings.

Publication Type: Article
Publisher Keywords: new software; automated counting; glistenings; intraocular lenses; slit-lamp images
Subjects: Q Science > QA Mathematics > QA75 Electronic computers. Computer science
R Medicine > RE Ophthalmology
Departments: School of Health & Psychological Sciences > Optometry & Visual Sciences
Professional Services
SWORD Depositor:
[thumbnail of A_new_software_for_automated_counting_of_glistenings_in_intraocular_lenses_in_vivo.pdf]
Preview
Text - Published Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

Download (1MB) | Preview
[thumbnail of A new software for automated counting of glistenings in intraocular lenses in vivo_060222_CLEAN_.pdf] Text - Accepted Version
This document is not freely accessible due to copyright restrictions.

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login