City Research Online

Optimisation of an Electrical Impedance Sensor for Use in Microfluidic Chip Electrophoresis

Hantschke, M. ORCID: 0000-0002-2664-0829 & Triantis, I. F. ORCID: 0000-0002-8900-781X (2022). Optimisation of an Electrical Impedance Sensor for Use in Microfluidic Chip Electrophoresis. IEEE Sensors Journal, 22(1), pp. 16-24. doi: 10.1109/jsen.2021.3127320

Abstract

Label-free measurements using impedance sensing could enable microfluidic chip electrophoresis (ME) to be used in point-of-care (POC) diagnostics. However, impedance sensing methods reported in this field need considerable optimisation. GOAL: Develop a novel design process for optimising tetrapolar electrical impedance measurement (TEIM) sensor performance in ME applications through a systematic investigation a) of the impact of a TEIM sensor's design parameters on its performance in ME and b) of the relationship between the above parameters with those of the microfluidic channel and the sample to be sensed. METHODS: 3D FEM sensitivity simulations, verified experimentally, were carried out to study the impact of sensor and channel parameters on the measured impedance and their interrelationship. Subsequently, the impact of sensor parameters on sensing a sample band's conductivity and size was investigated. RESULTS: The impact of channel dimensions on transfer impedance measurements is significant. The non-linearity reported for transfer impedance measurement of volume conductors can be manipulated by appropriate sensor parameter design. The sensor performance can be optimised by designing electrode length and measurement electrode distance in relation to the channel height and sample band length, respectively. The sensor performance is not affected by the injection electrode distance. CONCLUSION: There is a relationship between sensor, channel and band parameters and this warrants establishing a systematic design process of TEIM sensors in ME. SIGNIFICANCE: This paper presents a novel approach to optimising the design of TEIM sensors in ME potentially providing significant performance improvements and thus allowing for label-free POC electrophoresis diagnostics.

Publication Type: Article
Additional Information: © 2022 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
Publisher Keywords: Tetrapolar, Impedance, Sensitivity, Microfluidics, Electrophoresis, Cell constant, Conductivity Detection
Subjects: Q Science > QA Mathematics > QA75 Electronic computers. Computer science
T Technology > TK Electrical engineering. Electronics Nuclear engineering
Departments: School of Science & Technology
School of Science & Technology > Engineering
SWORD Depositor:
[thumbnail of Optimisation_of_an_Electrical_Impedance_Sensor_for_use_in_Microfluidic_Chip_Electrophoresis.pdf]
Preview
Text - Accepted Version
Download (1MB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login