City Research Online

Reliability of Self-Monitoring of Intraocular Pressure With iCare Home2 Rebound Tonometry

Romano, D., Montesano, G. ORCID: 0000-0002-9148-2804, Aminoleslami, A. A. , Colizzi, B. & Rossetti, L. M. (2025). Reliability of Self-Monitoring of Intraocular Pressure With iCare Home2 Rebound Tonometry. Journal of Glaucoma, 34(6), pp. 447-454. doi: 10.1097/ijg.0000000000002560

Abstract

Précis:
Using iCare Home2 (iCare, Finland) rebound tonometry, self-measurement of intraocular pressure has demonstrated good reliability and ease of use.

Purpose:
To investigate the reliability and repeatability of self-measured intraocular pressure (IOP) with rebound tonometry using iCare Home2.

Patients and Methods:
One hundred four patients out of 110 consecutive patients were recruited for this observational cross-sectional study. One randomly selected eye from each patient underwent 6 consecutive IOP measurements with Goldmann applanation tonometry (GAT), iCare IC200, and iCare Home2. Every eye was tested twice with each device, in random order, by an ophthalmologist for GAT and IC200, and by the patient itself for Home2. In addition, central corneal thickness (CCT) has been collected. The reliability of Home2 has been tested by calculating limits of agreement (LoA) between self-measured and physician-measured IOP, using the Bland-Altman analysis. The repeatability of each device has been tested by calculating the limits of repeatability (LoR) with the same method. Pearson correlation coefficient was used to determine the correlation between differences in IOP measurements and CCT.

Results:
The mean difference between GAT and iCare Home2 was −0.28±1.57 mmHg (P=0.070), 95%-LoA: (−3.36 to 2.79 mmHg). The mean difference between IC200 and iCare Home2 was 0.92±1.48 mmHg (P<0.0001), 95%-LoA (−1.98 to 3.82 mmHg). The mean difference between the first and second measurements with GAT, iCare IC200, and iCare Home2 measurements was 0.21±0.98 mmHg (P=0.03), −0.02±1.11 mmHg (P=0.88) and −0.23±1.04 mmHg (P=0.05).

Conclusions:
Self-measured IOP with rebound tonometry showed good reliability and repeatability when compared with physician-measured IOP with both standard GAT and rebound tonometry.

Publication Type: Article
Additional Information: This is a open access article distributed under the Creative Commons Attribution License 4.0. Copyright © 2025 The Author(s). Published by Wolters Kluwer Health, Inc.
Publisher Keywords: glaucoma, intraocular pressure, self-monitoring, home-monitoring, rebound tonometry
Subjects: R Medicine > RE Ophthalmology
Departments: School of Health & Medical Sciences
School of Health & Medical Sciences > Department of Optometry & Visual Science
SWORD Depositor:
[thumbnail of reliability_of_self_monitoring_of_intraocular.5.pdf]
Preview
Text - Published Version
Available under License Creative Commons: Attribution International Public License 4.0.

Download (596kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login