Purpose: To design and test a new noninvasive method for anterior chamber angle (ACA) estimation based on the slit lamp that is accessible to all eye-care professionals.

Methods: A new technique (slit lamp anterior chamber estimation [SLACE]) that aims to overcome some of the limitations of the van Herick procedure was designed. The technique, which only requires a slit lamp, was applied to estimate the ACA of 50 participants (100 eyes) using two different slit lamp models, and results were compared with gonioscopy as the clinical standard.

Results: The Spearman nonparametric correlation between ACA values as determined by gonioscopy and SLACE were 0.81 (p < 0.001) and 0.79 (p < 0.001) for each slit lamp. Sensitivity values of 100 and 87.5% and specificity values of 75 and 81.2%, depending on the slit lamp used, were obtained for the SLACE technique as compared with gonioscopy (Spaeth classification).

Conclusions: The SLACE technique, when compared with gonioscopy, displayed good accuracy in the detection of narrow angles, and it may be useful for eye-care clinicians without access to expensive alternative equipment or those who cannot perform gonioscopy because of legal constraints regarding the use of diagnostic drugs.

Publication Type: Article
Additional Information: This is a non-final version of an article published in final form in Optometry and Vision Science Issue: Volume 91(6), June 2014, p 668–675.
Publisher Keywords: angle closure, anterior chamber angle, gonioscopy, limbal chamber depth, narrow angle glaucoma
Subjects: R Medicine > RE Ophthalmology
Departments: School of Health & Psychological Sciences > Optometry & Visual Sciences
SWORD Depositor:
[thumbnail of A NEW SLIT LAMP BASED TECHNIQUE FOR ANTERIOR CHAMBER ANGLE ESTIMATION.pdf]
Preview
PDF
Download (214kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login