The effect of videokeratoscope faceplate design on radius of curvature maps

Monteiro, P.M.L. & Hull, C.C. (2007). The effect of videokeratoscope faceplate design on radius of curvature maps. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 27(1), pp. 76-84. doi: 10.1111/j.1475-1313.2006.00426.x

Abstract

A computer model using finite ray tracing methods was developed to simulate a videokeratoscope analysing an average cornea. Different faceplate designs were tested using five points in the faceplate subtending angles between 15 and 75 in 15 intervals at the corneal vertex. Image quality was assessed by adding the geometrical blurs of the 5 image points. Differences (error) between accurate sagittal radius of curvature and sagittal radius of curvature calculated by the van Saarloos algorithm were calculated for selected surfaces at the same corneal points. The calculations were repeated for the tangential radius of curvature. Differences equal or bigger than 0.02 mm were regarded as clinically significant. The surface that provided the sharpest image for an average cornea was a cylinder with the base 120 mm away from the corneal vertex and a diameter of 26 mm. Changing the faceplate design results in clinically significant differences for an average cornea.

Publication Type: Article
Publisher Keywords: Videokeratoscopy; Corneal topography; Faceplate design; Radius of curvature maps
Subjects: R Medicine > R Medicine (General)
Departments: School of Health & Psychological Sciences > Optometry & Visual Sciences
SWORD Depositor:
[thumbnail of Effect_of_Faceplate_design_on_radius_of_curvature_maps-_final_author_MS.pdf]
Preview
PDF - Accepted Version
Download (310kB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login