Repeatability and agreement of central corneal thickness measurements with a new handheld non-contact pachymeter

Lawrenson, J. ORCID: 0000-0002-2031-6390, Gill, S., Masuid, I. & Rashid, F. (2023). Repeatability and agreement of central corneal thickness measurements with a new handheld non-contact pachymeter. Ophthalmic and Physiological Optics, 43(6), pp. 1524-1530. doi: 10.1111/opo.13199

Abstract

Purpose
To compare the repeatability of central corneal thickness (CCT) measurements taken with a new handheld pachymeter (Occuity PM1 pachymeter) and to assess its agreement with ultrasound biometry and two commercially available optical biometers in participants with normal eyes.

Methods
Three consecutive CCT measurements of the right eye of 105 participants with normal corneas were acquired by the PM1 pachymeter, Lenstar LS 900 and Oculus Pentacam HR in a random order. This was followed by three measurements with a handheld ultrasound pachymeter (UP) (Pachmate 2). Repeatability and the repeatability limit were calculated for each device and Bland–Altman limits of agreement (LoA) were determined for the PM1 pachymeter compared to the other devices.

Results
The mean CCT (±SD) was 551.04 ± 33.43, 558.62 ± 31.46, 549.41 ± 31.00 and 539.73 ± 29.50 μm for the PM1 pachymeter, UP, Lenstar and Pentacam, respectively. The repeatability limits (expressed as the within subject SD for repeat measurements) were 14.02, 13.68, 4.99 and 9.90 μm, respectively. The closest agreement was between the PM1 and Lenstar (mean difference = −1.63 μm with LoA 10.72 μm below and 13.97 μm above the readings obtained with the Lenstar). The PM1 underestimated CCT compared to UP (mean difference = 7.58 μm, LoA 24.63 μm below and 9.47 μm above UP). The agreement was lowest between the PM1 and Pentacam (mean difference = −11.30 μm, LoA between 4.29 and 26.89 μm).

Conclusions
The PM1 pachymeter shows excellent precision for CCT measurements across a range of corneal thicknesses in normal eyes and provides a safe and easy-to-use alternative to ultrasound pachymetry.

Publication Type: Article
Additional Information: This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. © 2023 The Authors. Ophthalmic and Physiological Optics published by John Wiley & Sons Ltd on behalf of College of Optometrists.
Publisher Keywords: agreement, central corneal thickness, pachymetry, precision
Subjects: R Medicine > RE Ophthalmology
Departments: School of Health & Psychological Sciences > Optometry & Visual Sciences
SWORD Depositor:
[thumbnail of Ophthalmic Physiologic Optic - 2023 - Lawrenson - Repeatability and agreement of central corneal thickness measurements.pdf]
Preview
Text - Published Version
Available under License Creative Commons: Attribution International Public License 4.0.

Download (1MB) | Preview

Export

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Downloads

Downloads per month over past year

View more statistics

Actions (login required)

Admin Login Admin Login